Search our store

£625.00 £750.00 (Inc. VAT)

999 in stock

Product Description

Hanna


Hanna HI-2631 Conductivity Benchtop meter with Data Logging

Product code: HI-2631-02

HI-2631 benchtop meter enables fast, accurate measurements of conductivity, TDS, salinity, and temperature utilising Hanna digital probes – HI-763100 Digital Conductivity electrode with temperature sensor. HI-2631 also adds salinity measurement and calibration along with data logging capabilities. Ideal for laboratories, classrooms and general industry.

The HI-2631 architecture brings together Hanna’s visionary design capabilities, integrated production, and world class research and development.

  •  Conductivity & TDS
  • Salinity

HI-2631 benchtop meter enables fast, accurate measurements utilising the Hanna Instruments digital electrode for conductivity (HI-763100) which includes an integral temperature sensor. Each digital electrode has a unique serial number and the parameter that it measures is automatically identified by the meter.

The 2631 measures conductivity and TDS while adding salinity measurement and calibration along with data logging capabilities.

The intuitive design simplifies configuration, calibration, measurement, data logging and transfer of data to a USB thumb drive or computer.

The benchtop meter offers a Basic Mode that streamlines measurement configuration and is useful for routine applications. Additionally, the HI-2631 can be used in Standard Operating Mode, where all features and capabilities are enabled.

The meter has a large display with capacitive keys which features a 5.5” (14 cm) LCD. The large display provides a 130 ° wide viewing angle.

 

Main feature points

  • Automatic parameter recognition
  • pH/ORP choice of measurement units: pH, mV;
  • CAL Check indicators
  • Sensor Check indicators
  • Basic mode for simplified operation
  • Dedicated GLP key
  • GLP data included with logged data
  • Simplified data transfer to a PC
  • Internal clock and date

Key Features

Digital probe input – simply connect each probe via the 3.5 mm jack, Digital Smart Electrodes are automatically recognised.

Clear, full text readout – The HI-2631 presents clear, detailed text prompts at the bottom section of the screen.

Two operating modes – HI-2631 can be used in Standard or Basic Operating Modes. Standard Mode enables all features while Basic Mode displays a simplified screen and features which is ideal for routine measurements.

Data logging – The HI-2631 provides ample data storage capacity, allowing you to save up to 1000 log records. Each data set comprehensively includes readings, GLP information, as well as date and time stamps.

GLP – The sensor retains data from the most recent calibration, including date, time, and buffers utilised. When connected to the HI-2631, this GLP information is automatically transferred.

Capacitive touch keypad – HI-2631 features sensitive capacitive touch buttons for accurate keystrokes for navigating menus and screens. Since they are part of the screen, the buttons can never get clogged with sample residue.

Easy to read LCD – HI-2631 features a 5.5” (14 cm) LCD display that you can clearly view from over 5 m (16.4’). The large display, with its wide 130° viewing angle, provides one of the easiest to read LCDs in the industry.

USB-C port for charging and data transfer – The HI-2631 features a streamlined USB-C port that facilitates charging the integrated rechargeable battery, transferring data to your computer, and exporting files directly to a flash drive.

 

What’s included:

  • HI-763100 digital conductivity electrode
  • HI-764026 Electrode holder with the following accessories: base plate with integrated pivot pin; cable holder clip, attached
  • 12880 µS/cm calibration solution, x 2 sachets
  • 1413 µS/cm calibration solution, x 4 sachets
  • 5000 µS/cm calibration solution, x 2 sachets
  • Electrode rinse solution x 2 sachets
  • HI-920018 – 1.5m USB-C to USB-C cable
  • USB-C power adapter
  • Instrument quality certificate
  • Probe / electrode quality certificate
  • Battery safety insert
  • Quick reference guide with QR code for instruction manual download

Warranty – Meter has 2 years. 6 months from date of purchase for electrodes and probes

 

More accessories to buy separately

Electrodes

  • HI-763100 Digital EC / temperature probe

 

More solutions to buy separately

Conductivity solutions

  • HI-7030L 12880 µS/cm calibration solution, 500 mL
  • HI-7030M 12880 µS/cm calibration solution, 230 mL
  • HI-7031L 1413 µS/cm calibration solution, 500 mL
  • HI-7031M 1413 µS/cm calibration solution, 230 mL
  • HI-7033L 84 µS/cm calibration solution, 500 mL
  • HI-7033M 84 µS/cm calibration solution, 230 mL
  • HI-7034L 80000 µS/cm calibration solution, 500 mL
  • HI-7034M 80000 µS/cm calibration solution, 230 mL
  • HI-7035L 111800 µS/cm calibration solution, 500 mL
  • HI-7035M 111800 µS/cm calibration solution, 230 mL
  • HI-7039L 5000 µS/cm calibration solution, 500 mL
  • HI-7039M 5000 µS/cm calibration solution, 230 mL
  • HI-7037L 100% NaCl sea water standard solution, 500 mL

 

Other useful accessories

  • HI-764026 Electrode holder for HI-2600 family

 

Models in the series

HI-2600 pH, EC, DO Benchtop Meter with Data Logging

HI-2620 pH and ORP Benchtop Meter

HI-2621 pH and ORP Benchtop Meter with Data Logging

HI-2630 EC & TDS Benchtop Meter

HI-2631 EC, TDS and Salinity Benchtop Meter with Data Logging

HI-2640 Dissolved Oxygen Benchtop Meter

HI-2641 Dissolved Oxygen Benchtop Meter with Data Logging

 

EC – using HI-763100 four-ring conductivity probe
   EC Range 0.00 to 29.99 µS/cm;
30.0 to 299.9 µS/cm;
300 to 2999 µS/cm;
3.00 to 29.99 mS/cm;
30.0 to 200.0 mS/cm;
up to 500.0 mS/cm,
and absolute conductivity (TDS) is the conductivity value without temperature compensation
 Resolution 0.01 µS/cm;
0.1 µS/cm;
1 µS/cm;
0.01 mS/cm;
0.1 mS/cm
 Accuracy (@25°C) ±1% of reading (±0.5 µS/cm or 1 digit, whichever is greater)
 Calibration Cell factor calibration; six standards available: 84µS/cm, 1413 µS/cm, 5.00 mS/cm, 12.88 mS/cm, 80.0 mS/cm, 118.8 mS/cm, one point offset: 0.00 µS/cm
   Temperature Compensation Automatic, -5 to 100.0°C; NoTC, can be selected to measure absolute conductivity
 Temperature Coefficient 0.00 to 6.00%/°C (for EC and TDS only), default value is 1.90%/°C
TDS – using HI-763100 four-ring conductivity probe
   TDS Range 0.00 to 14.99 mg/L (ppm);
15.0 to 149.9 mg/L (ppm);
150 to 1499 mg/L (ppm);
1.50 to 14.99 g/L;
15.0 to 100.0 g/L;
up to 400.0 g/L absolute TDS using 0.80 conversion factor
 Resolution 0.01 mg/L (ppm);
0.1 mg/L (ppm);
1 (ppm);
0.01 g/L;
0.1 g/L
 Accuracy (@25°C) ±1% of reading (±0.03 ppm or 1 digit, whichever is greater)
 Calibration through EC calibration
 TDS Factor in Standard Mode 0.40 to 0.80 (default value is 0.50)
Salinity – using HI-763100 four-ring conductivity probe
   Salinity Range 0.0 to 400.0 % NaCl;
2.00 to 42.00 PSU;
0.00 to 80.00 g/L
 Resolution 0.1 % NaCl;
0.01 PSU;
0.01 g/L
 Accuracy (@25°C) ±1% of reading
 Calibration % NaCl only one point with HI-7037 standard
Temperature
   Temperature Range -20.0 to 120.0 °C
 Resolution 0.1°C
 Accuracy ±0.5°C
Additional Specifications
Logging Up to 1000 (standard mode) (400) records organised in:
• Manual log-on demand, maximum 200 logs
• Manual log-on-stability, maximum 200 logs
• Interval logging in Standard Mode, maximum 600 logs (100 lots)
PC Interface USB-C
Storage Interface USB-C
Power supply USB Type C (5 VDC; 500mA)
Environment 0 to 50˚C; Maximum 95% RH non-condensing
Dimensions 205 x 160 x 77 mm
Weight Approximately 0.85 kg